EMC Tables
28
Guidance and manufacturer’s declaration – electromagnetic emissions
The product is intended for use in the electromagne c environment specified below. The customer or the user of the
product should assure that it is used in such an environment.
Emissions
Compliance
Electromagnetic environment-- guidance
RF emissions
CISPR 11
Group 1
The product uses RF energy only for its internal func on. Therefore,
its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class A
The product is suitable for use in all establishments, including
domes c establishments and those directly connected to the public
low-voltage power supply network that supplies buildings used for
domes c purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctua ons/
flicker emissions
IEC 61000 -3-3
Complies
Guidance and manufacturer’s declaration 211; electromagnetic immunity
The product is intended for use in the electromagne c environment specified below. The customer or the user of the product
should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment guidance
Electrosta c
discharge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be wood, concrete or ceramic
le. If floors are covered with synthe c
material, the rela ve humidity should be at
least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±2 kV for power
supply lines
Mains power quality should be that of a
typical commercial or hospital environment.
Surge
IEC 61000-4-5
±1 kV line(s) and neutral
±1 kV line(s) and neutral
Mains power quality should be that of a
typical commercial or hospital environment.
Voltage dips, short
interrup ons and
voltage varia ons
on power supply
input lines
IEC 61000-4-11
<5 % U
T
(>95 % dip in U
T
)
for 0,5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
(>95 % dip in U
T
)
for 5s
<5 % U
T
(>95 % dip in U
T
)
for 0,5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles)
<5 % U
T
(>95 % dip in U
T
) for 5s
Mains power quality should be that of a
typical commercial or hospital environment.
If a dips or an interrup on of mains power
occurs, the current of the product may be
dropped off from normal level, it may be
necessary to use uninterrup ble power
supply or a ba ery.
Power frequency
(50/60 Hz)
magne c field
IEC 61000-4-8
3 A/m
30 A/m
Power frequency magnetic fields should be
at levels characteristic of a typical location in
a typical commercial or hospital
environment.
NOTE U
T
is the a.c. mains voltage prior to applica on of the test level