MD 11
MD 11, V2.0, Operating Instructions, REF 31665, V22/17
17
EN
7.8
Configuration
Menu
In the Configuration Menu the user can customize the device after his favor. The parameters are organized
in two levels. The following information and parameters are stored in there.
• Software version
• Mainboard serial number
• Language selection, DE/EN
• Display illumination
• Number of programs
• Power on, at last program
• Operating hours MD 11
• Operating hours Motor
• Operating hours of irrigation pump
• Error log
• Activation of available handpiece
• Speed limitation for each handpiece
• Range of action of AS mode
• Pump behavior
• Motor behavior
• Behavior of torque at reverse mode
• Reset to factory default
1.
Access to Configuration Menu:
• Press
“HP”
and
“Speed
▼
”
key for about 3 sec. until you hear a tone.
On the display the first position of the configuration menu shows up.
2.
Selection of parameters:
• Press
“HP”
or
“PR”
to choose the desired parameter.
• Press
“Speed
▼
”
to select the parameter.
• Change the value by pressing
“HP” for up
or
“PR” for down.
• To confirm the changes made, press
“Speed
▼
”
for about one sec
., until you hear a short tone.
• To abort the settings made, press
“Speed
▼
”
briefly. The setting returns to its previous value.
3.
Exit from Configuration Menu:
• To leave the Configuration Menu press
“HP” and “PR”
for 3 sec.
until you hear a tone.
Be cautious when changing parameters. Unusual behavior of instruments while operating may provoke
false reactions and jeopardize the patient.
Every setting and the new behavior of the instrument has to be verified and tested.
Parameter Level 1
Group/Parameter Rights
Factory
Definition
Software/Version
read
VX.XXXX
Shows current software version
Hardware/Serial number MB
read
XXXXXXXXXXX Shows serial number of the main board
Language, 0 = DE, 1 = EN
read/write
1
Select language between German and English
Backlight/brightness (0. .10)
read/write
9
Display brightness, changeable: 0, … , 10
Number of programs
read/write
4
Select between 2 and 6 available programs
Power on at last program
read/write
Yes
Display of the last program after switching device on.
Operating hours/Control Unit
read
0
Shows operating hours of Control Unit
Operating hours/Motor
read
0
Shows operating hours of motor 1
Operating hours/Pump
read
0
Shows operating hours of pump
Error memory/ 1 – 8
read
0
8 Error messages in chronological order.
+
The arrows at the beginning of the lines indicate
the operation in the Configuration Menu.
Содержание MD11
Страница 2: ......
Страница 4: ......
Страница 36: ......
Страница 68: ......
Страница 100: ......
Страница 132: ......
Страница 164: ......
Страница 196: ......
Страница 228: ......
Страница 229: ...EN DE FR IT ES Anhang Appendix Appendice Appendice Apéndice ...
Страница 230: ......
Страница 231: ......
Страница 234: ......