MD 11
MD 11, V2.0, Operating Instructions, REF 31665, V22/17
19
EN
Exit from Configuration Menu
• To leave the Configuration Menu press
“HP”
and
“PR”
for 3 sec., until you hear a long tone.
Motor type 21, 40,000 rpm
Range
Factory
default
Definition
Motor type 21, 40,000 rpm/Min. speed 300 – 5000 rpm
300 rpm
Set the min. speed of the motor.
Motor type 21, 40,000 rpm/Max. speed 5000 – 40,000 rpm
40,000 rpm Set max. speed of the motor.
Motor type 21, 40,000 rpm /Start ramp 1 – 1000 ms/10,000 rpm 100 ms
Set acceleration time to 10,000 rpm
Motor type 21, 40,000 rpm /Stop ramp 1 – 1000 ms/10,000 rpm 50 ms
Set breaking time from 10,000 – 0 rpm
The maximum torque of the motor is boosted while starting the reverse rotation to disengage stuck
implants or screws.
Reverse torque
Range
Factory
default
Definition
Reverse torque/Increase
5 – 30 %
25 %
Increase of the selected torque in reverse
rotation.
Reverse torque/Increase time
100 – 2000 ms
500 ms
Time during which torque is raised.
•
Attention:
Upon resetting to factory default all parameters will appear with factory default values
(except date and time).
Resetting to
factory default
Choice Factory
default
Definition
Default value/Set default value
Yes/No
–
Resetting all parameters of the Configuration
Menu to factory default.
Содержание MD11
Страница 2: ......
Страница 4: ......
Страница 36: ......
Страница 68: ......
Страница 100: ......
Страница 132: ......
Страница 164: ......
Страница 196: ......
Страница 228: ......
Страница 229: ...EN DE FR IT ES Anhang Appendix Appendice Appendice Apéndice ...
Страница 230: ......
Страница 231: ......
Страница 234: ......