![Northern Meditec Taurus E Скачать руководство пользователя страница 85](http://html1.mh-extra.com/html/northern-meditec/taurus-e/taurus-e_operators-manual_1708589085.webp)
文件编号:
NR-TP-0715-008
版本:
1.0
- 73 -
To prevent damage to the monitor, do not disinfect the monitor with gas (EtO) or
formaldehyde.
14.4
Sterilization
In order to avoid long-term damage to the device, we recommend that the product is
sterilized only when it is deemed necessary by the instrument maintenance program. We also
recommend that the instrument to be sterilized must first be cleaned.
Sterilization materials recommended for the monitor: ethanol group, aldehyde group.
Caution
Dilute or use concentration as low as possible according to the manufacturer's
instructions.
Do not let liquid enter into the enclosure.
Never immerse any part of the instrument.
Do not pour liquid onto the instrument in the sterilization process.
Do not leave any fungicide residues on the surface of the device; clean it
immediately with a damp cloth.