7.3.4
Parameter Flashing
When a physiological parameter of the patient alarms, the parameter values in the parameter
area will flash once per second, and the upper limit and lower limit of the parameter will also
flash at the same frequency, indicating that the parameter exceeds the upper limit or lower
limit.
7.4
Alarm States
In addition to the above alarm modes, you can also set the monitor to the following four
alarm states as needed, and display different alarm icons on the screen:
Alarm silence
Alarm sound off
Alarm pause
Alarm off
7.4.1
Silence
Select
[Silence] smart hotkey, and you can temporarily turn off the alarm sound of
currently occurring physiological alarms of the monitor, but the alarm information is still
retained. For technical alarms, clear the alarm state, display alarm prompt information only
and the monitor enters alarm silence state, and the alarm state icon area displays the
icon. When a new physiological alarm or technical alarm occurs, the alarm silence is
automatically canceled.
7.4.2
Alarm sound off
The alarm sound can be turned off through the following operations:
Select
[Menu] Smart Hotkey→[Main Menu];
Select [User Maintenance >>] →enter the password and confirm → [User Maintenance]
menu;
Select [Alarm Param >>] → [Ala.Para Setup] menu;