C.2
Technical Alarm Information
Source
Alarm message
L
Causes and countermeasures
System Battery Low
H
Connect to AC power supply, and charge the
battery, and power with the battery as needed after
fully charged.
ECG
ECG Comm. Stop
H
ECG module failure, or communication failure
between the module and the host; please restart
the device.
ECG Comm. Error
H
ECG Config Error
H
ECG Self check Error
H
ECG Lead Off
M*
The electrodes are not connected to the patient
firmly or fall off, or lead wires and the main cable
fall off. Check the connection of electrodes and
lead wires.
ECG YY OFF (YY is
a lead name)
M*
Temp
TEMP1 Sensor Off
L
The temperature sensor falls off from the patient.
Check the sensor connection.
TEMP2 Sensor Off
L
SpO2
SPO2 Comm. Stop
H
SpO2 module failure, or communication failure
between the module and the host; please restart
the device.
SPO2 Comm. Error
H
SPO2 No Sensor
L
SpO2 sensor falls off from the patient or monitor,
malfunctions, or sensor other than specified in this
Manual is used. Check the sensor mounting
position, whether the sensor is damaged or sensor
type. Reconnect the sensor or use new sensor.
SPO2 Sensor Off
L
SPO2 Sensor Error
L
SPO2 Search Timeout
L
SPO2 Search Pulse
L
Sensor signal is poor or too weak. Check the
patient's condition, and place the sensor in a
suitable position. If the failure persists, replace the
sensor.
SPO2 Signal Unstable
L
SPO2 Failure
L
SPO2 Signal Weak
L
NIBP
NIBP Comm. Stop
H
NIBP module failure, or communication failure
between the module and the host; please restart
the device.
NIBP Comm. Error
L
NIBP Self check error
H
NIBP CFG Error
H
NIBP system error
H
If failure occurs during measurement, the system
can’t analyze and calculate. Check the patient's
Measurement timeout
L