文件编号:
NR-TP-0715-008
版本:
1.0
- 88 -
Guidance and Declaration - Electromagnetic Immunity
Taurus
®
Patient Monitor is suitable for use in the electromagnetic environment specified below.
The customer or the user of the device should assure that it is used in such an environment.
Immunity test
IEC 60601 Test level
Compliance level
Conduced RF IEC61000-4-6
3 Vrms
150k to 80M Hz
3 Vrms
Radiated RF IEC61000-4-3
3V/m
80M to 2.5G Hz
3V/m
Electromagnetic environment - guidance
Portable and mobile RF communications equipment should be used no closer to any part of the
device, including cables, than the recommended separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended Separation Distance:
d
-
1.2
(
d
-
3.5
)
d
-
1.2
(Resp:
d
-
3.5
) 80 to 800MHz
d
-
1.2
800M to 2.5GHz
Where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer and d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, a
should be less than the compliance level in each frequency range b Interference may occur in the
vicinity of equipment marked with the following symbol:
Note 1: From 80 MHz to 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to
fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field
strength in the location in which the device is used exceeds the applicable RF compliance level
above, the device should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as reorienting or relocating the device.
b
Over the frequency ranges 150kHz to 80MHz, field strengths should be less than 3V/m.