Environmental Requirements
The storage, transport and use of the monitor must meet the following environmental
requirements.
Working
environment
Ambient
temperature range
5
℃
~40
℃
Relative
humidity
range
≤80%
Atmospheric
pressure range
86kPa~106kPa
Supply voltage
requirement
Supply voltage
a.c.100V~240V
Power frequency
50Hz/60Hz
Input power
40VA-60VA
Fuse
FUSE T 1.6A
Transportation
Prevent severe shock, vibration, rain and snow splashing during transport.
Storage
The packaged monitor should be stored in well-ventilated room with
ambient temperature -20
℃
~+40
℃
, relative humidity lower than 80%,
atmospheric pressure 50kPa~106kPa, and without corrosive gases.
The operating environment of the monitor should avoid noise, vibration, dust, corrosive or
flammable and explosive materials. In order to allow air flowing smoothly and achieve good
heat dissipation, at least 2 inches (5cm) clearance should be kept around the device.
When the device is moved from one environment to another, the device may have
condensation due to differences in temperature or humidity. In this case, wait until the
condensation disappears before using the device.
Warning
Please ensure that the device operates under stipulated environment, or else
technical specifications declared in this Manual may not be met, and it may
result in damage to equipment and other unforeseen consequences.
2.3
Connecting to AC Power
Connect to AC power in the following steps: