As shown in 6.4.3, if the [Alarm Pause Time] is set to [Permanent], press the
[Pause]
smart hotkey, and the monitor will turn off the alarm. In this case, except the alarm prompt
characteristics maintained in alarm pause state:
Alarm state icon area will display the
icon.
The physiological alarm area displays [Alarm Pause].
You can press the
[Pause] smart hotkey again to manually cancel the alarm off.
If the monitor is in the alarm state of suspension or closure of senior technical alarm is
triggered, the alarm and the alarm off pause are automatically canceled.
Warning
When the alarm volume is set to 0 or the alarm pause time is set to permanent,
the monitor does not sound an alarm when an alarm occurs. Therefore, the
operator should use this feature carefully.
7.5
Alarm Setup
7.5.1
Setting the Alarm Sound
See 3.9.4 Volume Control for the method to set the alarm volume.
7.5.2
Setting the Alarm Delay Time
For over-limit alarm of continuous measurement parameter, you can set the alarm delay time.
If the alarm condition disappears during the delay period, the monitor won’t alarm. In
[Ala.Param Setup] menu, select [Alarm Delay] time and [ST Alarm Delay] time.
The specific operation is as follows:
Select
[Menu] Smart Hotkey→[Main Menu];
Select [User Maintenance >>] →enter the password and confirm → [User Maintenance]
menu;
Select [Alarm Param >>] → [Ala.Param Setup] menu;