Integrator's Guide
26
© 2017 Nortek AS
5.6
SETAVG/GETAVG/GETAVGLIM
Set/get averaging mode settings and get the relevant limits.
Argument
Description
NC
Number of cells
CS
Cell Size (m)
BD
Blanking Distance (m)
CY
Coordinate System (“ENU”, ”XYZ“, “BEAM”)
PL
Power Level [dB]
(-100 dB to switch off transmit
-20.0 dB to 0.0 dB)
AI
Average interval (s)
VP
Unused
VR
Velocity range along beam [m/s]
DF
Data Format
0 – AD2CP format 1.
1 –Legacy Aquapro format
2 - AD2CP format 2
3 - AD2CP format 3
NPING
Number of pings
NB
Number of beams
(Select number of beams, 0 select beams according the PLAN,VD setting)
CH
Beam selection
(Select beams, 0 select beams according the PLAN,VD setting. Example: 134
select the three beams 1, 3 and 4)
MUX
Multiplexor Selection
0 - 1,
0, ping all beams in parallel
1, ping beams in sequence
BW
Bandwidth selection. (“NARROW”, “BROAD”)
ALTI
Enable altimeter in AVG measurements
BT
Enable bottom track measurement in AVG sampling
ICE
Enabled ice velocity measurement in AVG sampling
ALTISTART
Altimeter start of measurement distance [m]
ALTIEND
Altimeter end of measurement distance [m]
RAWALTI
Storage of Raw Altimeter data
The actual valid range for the various parameters for the firmware version is used can be found by
using the
GETAVGLIM
command. This command has the same arguments as the
SETAVG
/
GETAVG
commands shown in the list above. The output format for limits is described in
Содержание Signature 100