84
Chapter 18
Technical specifications
85
Norsonic Nor118
User Guide
Filter bands:
1⁄3 oct: 6.3 Hz to 250 Hz:
15 dB (option),
1⁄3 oct: 315 Hz to 20 kHz:
10 dB (option)
Noise measured with the input terminal on the sound
level meter short-circuited to ground, averaged over 30 s
of measurement time:
Spectral weighting functions:
A-weighted: 10 dB,
C-weighted: 10 dB, Z-weighted: 15 dB
Filter bands:
1⁄3 oct: 6.3 Hz to 250 Hz:
0 dB (option)
1⁄3 oct: 315 Hz to 20 kHz:
7 dB (option)
Field calibration
e recommended sound calibrator for verification of the
sensitivity of the sound level meter is Norsonic Nor1251
with a nominal sound pressure 114.0 dB at 1 kHz. In order
to compensate for effects due to diffraction around the
microphone, we recommend adjusting the sound level
meter to indicate 113.8 dB (diffuse correction off).
If other types of calibrators or electrostatic actuators
are used for the calibration, we recommend adjusting the
sound level meter to indicate the following levels referred
to the sound pressure acting on the diaphragm of the
microphone (diffuse correction off):
Freq
125 Hz
250 Hz
1 kHz
4 kHz 8 kHz
Level
0.0 dB
0.0 dB –0.2 dB –0.8 dB –2.8 dB
Measurement duration and resolution
e total time period for a measurement (global period,
designated
Σ
) may be set from 1 second up to 100 hours
less 1 second with 1 second resolution. e global period
may be subdivided in shorter periods, designated time
resolution (
∆
) from 1 second up to the global period. As
an option the time resolution may be set in an additional
range: from 100 millisecond and upwards to 1 second in
steps of 25 millisecond.
Total range for measurement of A-weighted levels
e linear operating range is identical to the total range.
Frequency 31.5 Hz
1 kHz
4 kHz
8 kHz 12.5 kHz
Upper level
98 dB 137 dB 138 dB 136 dB 133 dB
Lower level
24 dB
24 dB
24 dB
24 dB
24 dB
Ref level test
94 dB 114 dB
114 dB 114 dB
114 dB
e primary indicator range for compliance with
type 1 is 24 dB to 117 dB. For compliance with
type 1, the linearity range is 24 to 137 dB, and the pulse
range 24 dB to 140 dB, respectively.
Total range for measurement of C-weighted levels
e linear operating range is identical to the total range.
Frequency 31.5 Hz
1 kHz
4 kHz
8 kHz 12.5 kHz
Upper level
134 dB 137 dB 136 dB 134 dB 131 dB
Lower level
30 dB
30 dB
30 dB
30 dB
30 dB
Ref level test
114 dB 114 dB
114 dB 114 dB
114 dB
Total range for measurement of Z-weighted levels
e linear operating range is identical to the total range.
Frequency 31.5 Hz
1 kHz
4 kHz
8 kHz 12.5 kHz
Upper level
137 dB 137 dB 137 dB 137 dB 137 dB
Lower level
40 dB
40 dB
40 dB
40 dB
40 dB
Ref level test
114 dB 114 dB
114 dB 114 dB
114 dB