Chapter 10
Reverberation time measurements
57
Norsonic Nor118
User Guide
reverberation time stated in the table a few pages back a
question mark will be shown to the right of the value.
Likewise, the question mark will appear if the back-
ground noise level is too high. All in all, the question mark
tells you that the measurement is not reliable. However, it
does not tell you why it is so.
Overload occurring during the impulse decay capture
is indicated by an
OL
appearing above the result table. In
addition, no reverberation time will be calculated and the
table will reflect this.
Saving the measurement
While still displaying the result table the acquired data
may easily be stored.
To store the acquired data:
• Press to save the measurement. e measure-
ment will be stored in a folder bearing today’s date as
its name and the file name will be the first free number.
Existing files will not be over written. If the folder did
not exist, it will be created when you store the data.
Recalling a stored measurement
A stored measurement is retrieved as any other measure-
ment stored. e values will be presented in a table identi-
cal to the result table. See the chapter
Memory handling
for
a discussion of storage and retrieval of measured data.
Making another measurement
If you want to make another measurement, you may do
this with or without storing the one you just did.
To make another measurement:
• While in the result table, press the t key. You will
now be prompted to either store the acquired data (by
pressing ) or start another measurement directly
(by pressing again).
Leaving the table returning to
READY
mode
You may leave the table to go back to ready mode without
having to start another measurement first. is may be
the case if the trigger settings needs adjustment.
To leave the table:
• Press the key to leave the table. You will now be
prompted to either store the acquired data (by pressing
) or exit directly (by pressing again).
To store a reverbera-
tion time measurement,
press the
STORE
key
whilst still displaying
the reverberation time
results.