Chapter 7
Enhanced time profile measurements
43
Norsonic Nor118
User Guide
If you’ve set up a profile resolution different from (i.e.
shorter than) the global measurement duration, the
instrument will log the time profile in addition to the
global measurement.
e enhanced profile offers a time resolution (period
length) down to 100 ms! Between 100 ms and 1 s the
resolution is adjustable in 25 ms steps. Above 1 s the step
size is 1 s as is the case for the basic profile.
Setting a period length better than a second
• Set the resolution to 1 s and press the key to enter
the ms area. Scroll down to the required setting or
use the as usual after the ms sign
appears. Note that if you keep the key (or the
key) depressed it will, after a while, speed up the
scrolling.
or
• Key in 59 s as the resolution and then press enter
followed by , the resolution will be set to 100 ms
immediately. Press or to adjust, or use the
to set the value directly. Step size
is 25 ms. If you key in a value between the valid settings
the value will be put to the nearest valid setting.
No back erase in profile pause mode.
e resumption of a paused measurement
will cause the ten seconds acquired immediately
before the pause to be erased. is applies to the
global measurement, but not to the profile.
Assume you have set up the measurement to also
include profile and that the measurement is running.
If you press , the global data acquisition will be
halted, but the profile will still acquire data! ese
data will be marked for pause to denote that they
were acquired in pause mode and that they do not
participate in the global level assessments.
When you later press again to resume, the
amount of periods marked as paused will be
expanded backwards in time to include periods
acquired in the time-span subject to back-erasure
in the global mode.
e reason why we designed it this way, was to
provide you with the complete overview – if you later
transfer the acquired data to your you will be able
to do calculations on any parts of the profile while at
the same time you’ll be able to see the intervals that
didn’t take part in the global assessments.
Consequently, the duration of the measurement
will seem ambigous. e global duration will be less
the pause and less any back-erase, while the profile
will have a duration including the pause length and
without back-erase.
The pause marker
The line denotes
periods are
marked
The effect of pressing
STOP
before
resuming.
If you terminate an ongoing
measurement prematurely by pressing the
key and later resume the measurement by pressing
/ an marker will be added to the period
within which the / key was pressed – i.e.
we mark out the first period after resumption.