Chapter 18
Technical specifications
93
Norsonic Nor118
User Guide
Information for conformance testing
Reference Sound Pressure Level:
114.0 dB re 20 µPa.
e reference frequency is 1000 Hz.
Reference Level Range:
e instrument has one level
range only.
Microphone Reference Point and Direction:
e
microphone reference point is the geometric centre of
the diaphragm of the microphone. The microphone
reference direction is from the microphone and along
the axis of rotational symmetry for the microphone and
preamplifier.
Battery voltage:
e instrument will automatically
switch off if the battery or external voltage is too low for
operation within the stated specifications. e max. bat-
tery voltage for conformance testing is 4 ×1.6 V = 6.4 V.
Electromagnetic Compatibility:
When the instrument
is tested for conformance to electromagnetic compatibility
requirements, the instrument should be in the measure-
ment mode, as this normally will generate the highest
levels of emissions. e highest susceptibility is normally
observed when the display faces the principal direction
of propagation for the electromagnetic field.
Содержание nor118