437D140 Rev. I
Installation and Setup
3-29
Note:
If the
Scan Speed
is too fast for the chosen
Resolution
, the following message will appear at
the beginning of a scan:
If this occurs, the
Scan Speed
must be reduced or the
Resolution
must be increased.
Scan Width: entered in centimeters.
Minimum for
Scout
and
Measure
scans is 2 x point resolution
Maximum for
Scout
and
Measure
scans is 255 x point resolution
Note:
If the
Scan Width
is too large for the chosen
Resolution
, the following message will appear at
the beginning of a scan:
If this occurs, the
Scan Width
must be reduced or the
Resolution
must be increased.
Scout Scan: when enabled, provides image to assist in defining the Measure scan region
Force initial scan on-axis: forces end point of initial scan to be on same Y-axis as the start point
Maximum degrees to snap final scan on-axis: This field defines the maximum degree of rotation of the
cursor box at the scan progress screen (following a scout scan) that will result in an on-axis scan. On-axis
scans are faster than off-axis scans.
Rotation of the cursor box to an angle greater than or equal to the
Maximum degrees
setting will result in
scan data collection at that angle. Rotation of the cursor box to an angle less than the
Maximum degrees
set-
ting will result in an on-axis scan.
On-axis scans are scans taken with the start and end points on the same y-axis.
If the
Maximum degrees to snap final scan on-axis
is set to 15, then
Force on Axis
will correct scout view dis-
play for scout scans that are up to 15 degrees off axis.
Scan Type Name: operator-defined; default is “Research” or “Small Subject”
Содержание Elite 437A150
Страница 1: ...DXA Bone Densitometer Operator s Guide Elite Model 437A150 Operator s Guide 437D140 Rev I...
Страница 22: ...This page intentionally left blank Operator s Guide 437D140 Rev I 1 8...
Страница 41: ...Operator s Guide 437D140 Rev I General Information 2 19 Figure 2 6 Page 2 of the Bone Exam Report...
Страница 45: ...Operator s Guide 437D140 Rev I General Information 2 23 Figure 2 10 Page 2 of the Patient Letter...
Страница 47: ...Operator s Guide 437D140 Rev I General Information 2 25 Figure 2 12 Page 2 of the Referral Letter...
Страница 49: ...Operator s Guide 437D140 Rev I General Information 2 27 Figure 2 14 Page 2 of the Ten Year Fracture Risk Report...
Страница 56: ...This page intentionally left blank Operator s Guide 437D140 Rev I 2 34...
Страница 66: ...11 Click Operator s Guide 437D140 Rev I 3 10 Installation and Setup...
Страница 108: ...This page intentionally left blank Operator s Guide 437D140 Rev I 3 52...
Страница 124: ...Sample Printed QA Results Report Operator s Guide 437D140 Rev I 4 16 Basic Operation...
Страница 137: ...Operator s Guide 437D140 Rev I Basic Operation 4 29...
Страница 165: ...Operator s Guide 437D140 Rev I Scanning Whole Body 5 23 Figure 5 6 Page 2 of the Bone Exam Report...
Страница 166: ...This page intentionally left blank Operator s Guide 437D140 Rev I 5 24...
Страница 176: ...Figure 6 2 Page 2 of the Bone Exam Report Operator s Guide 437D140 Rev I 6 10 Soft Tissue Composition...
Страница 178: ...Figure 6 4 Page 2 of the Body Composition Report Operator s Guide 437D140 Rev I 6 12 Soft Tissue Composition...
Страница 198: ...Operator s Guide 437D140 Rev I 7 20 Scanning AP Spine...
Страница 205: ...Operator s Guide 437D140 Rev I Scanning AP Spine 7 27 Figure 7 5 Page 2 of the Bone Exam Report...
Страница 206: ...This page intentionally left blank Operator s Guide 437D140 Rev I 7 28...
Страница 238: ...Figure 8 6 Page 2 of the Bone Exam Report Operator s Guide 437D140 Rev I 8 32 Scanning Hip...
Страница 289: ...Operator s Guide 437D140 Rev I Scanning Lateral Spine 10 29 Figure 10 6 Page 2 of the Bone Exam Report...
Страница 290: ...This page intentionally left blank Operator s Guide 437D140 Rev I 10 30...
Страница 301: ...Operator s Guide 437D140 Rev I Additional Techniques 11 11 6 The dialog box closes and the file is sent...
Страница 326: ...Operator s Guide 437D140 Rev I 11 36 Additional Techniques...
Страница 378: ...Figure 12 2 Page 1 of the Ten Year Fracture Risk Report Operator s Guide 437D140 Rev I 12 10 Ten Year Fracture Risk...
Страница 412: ...This page intentionally left blank Operator s Guide 437D140 Rev I 14 16...
Страница 428: ...This page intentionally left blank Operator s Guide 437D140 Rev I 16 10...
Страница 446: ...Label Locations continued Operator s Guide 437D140 Rev I 17 18 Technical Reference...
Страница 448: ...Scanner Labels continued Operator s Guide 437D140 Rev I 17 20 Technical Reference...
Страница 449: ...Operator s Guide 437D140 Rev I Technical Reference 17 21 Scanner Labels continued...
Страница 457: ...Operator s Guide 437D140 Rev I Reference Data Sets 18 7 Excerpt from Calcified Tissue International...
Страница 458: ...Excerpt from the Journal of Bone and Mineral Research Operator s Guide 437D140 Rev I 18 8 Reference Data Sets...
Страница 459: ...Operator s Guide 437D140 Rev I Reference Data Sets 18 9...
Страница 466: ...Operator s Guide 437D140 Rev I This page intentionally left blank...
Страница 484: ...Operator s Guide 437D140 Rev I This page intentionally left blank...