Calibration
Condition
Action
Alert: Calibration points were taken in
the wrong order.
When marking the points on the calibration standard, the points were
marked in the wrong order, or were not marked in the proper location on
the calibration standard. See "Daily Calibration Procedure" on page 4-6
for instructions on performing the calibration.
Alert: Process timed out before
enough counts were collected. The x-
ray source is not generating enough
x-rays or the beam is blocked.
This error may appear if there is a power problem inside the scanner, or if
the x-ray source has failed.
Contact Technical Support for assistance.
Calibration Diagnostics: Fail
Autoset PMT Gains
Reference Beam
X-ray Off
Shutter
Low Atten. Filter
Med. Atten. Filter
High Atten. Filter
Dynamic Filtration
Verify the calibration standard is placed in the correct calibration area.
Verify points 'A' and 'B' were marked in the correct order and on the 'X' in
the Plexiglas of the calibration standard.
Turn off the scanner and computer and repeat the calibration.
If the calibration fails a second time, print the errors and contact Tech-
nical Support.
TO PRINT: Go to Tools > Calibration Details > diagnost.trc tab. Then
click on
.
Calibration Status: Failed
Verify the calibration standard (QA Phantom) is placed in the correct cal-
ibration area.
Verify points 'A' and 'B' were marked in the correct order and on the 'X' in
the Plexiglas of the calibration standard.
Verify the Calibration Standard has not been dropped and broken.
Turn off the scanner and computer and turn back on. Repeat the cal-
ibration.
If the calibration fails a second time follow the instructions below to print
the errors and contact Technical Support.
TO PRINT: Go to Tools > Calibration Details > errors.trc tab. Then
click on
.
Table 16-4:
Calibration Faults
Operator's Guide
437D140 Rev. I
16-6
Troubleshooting
Содержание Elite 437A150
Страница 1: ...DXA Bone Densitometer Operator s Guide Elite Model 437A150 Operator s Guide 437D140 Rev I...
Страница 22: ...This page intentionally left blank Operator s Guide 437D140 Rev I 1 8...
Страница 41: ...Operator s Guide 437D140 Rev I General Information 2 19 Figure 2 6 Page 2 of the Bone Exam Report...
Страница 45: ...Operator s Guide 437D140 Rev I General Information 2 23 Figure 2 10 Page 2 of the Patient Letter...
Страница 47: ...Operator s Guide 437D140 Rev I General Information 2 25 Figure 2 12 Page 2 of the Referral Letter...
Страница 49: ...Operator s Guide 437D140 Rev I General Information 2 27 Figure 2 14 Page 2 of the Ten Year Fracture Risk Report...
Страница 56: ...This page intentionally left blank Operator s Guide 437D140 Rev I 2 34...
Страница 66: ...11 Click Operator s Guide 437D140 Rev I 3 10 Installation and Setup...
Страница 108: ...This page intentionally left blank Operator s Guide 437D140 Rev I 3 52...
Страница 124: ...Sample Printed QA Results Report Operator s Guide 437D140 Rev I 4 16 Basic Operation...
Страница 137: ...Operator s Guide 437D140 Rev I Basic Operation 4 29...
Страница 165: ...Operator s Guide 437D140 Rev I Scanning Whole Body 5 23 Figure 5 6 Page 2 of the Bone Exam Report...
Страница 166: ...This page intentionally left blank Operator s Guide 437D140 Rev I 5 24...
Страница 176: ...Figure 6 2 Page 2 of the Bone Exam Report Operator s Guide 437D140 Rev I 6 10 Soft Tissue Composition...
Страница 178: ...Figure 6 4 Page 2 of the Body Composition Report Operator s Guide 437D140 Rev I 6 12 Soft Tissue Composition...
Страница 198: ...Operator s Guide 437D140 Rev I 7 20 Scanning AP Spine...
Страница 205: ...Operator s Guide 437D140 Rev I Scanning AP Spine 7 27 Figure 7 5 Page 2 of the Bone Exam Report...
Страница 206: ...This page intentionally left blank Operator s Guide 437D140 Rev I 7 28...
Страница 238: ...Figure 8 6 Page 2 of the Bone Exam Report Operator s Guide 437D140 Rev I 8 32 Scanning Hip...
Страница 289: ...Operator s Guide 437D140 Rev I Scanning Lateral Spine 10 29 Figure 10 6 Page 2 of the Bone Exam Report...
Страница 290: ...This page intentionally left blank Operator s Guide 437D140 Rev I 10 30...
Страница 301: ...Operator s Guide 437D140 Rev I Additional Techniques 11 11 6 The dialog box closes and the file is sent...
Страница 326: ...Operator s Guide 437D140 Rev I 11 36 Additional Techniques...
Страница 378: ...Figure 12 2 Page 1 of the Ten Year Fracture Risk Report Operator s Guide 437D140 Rev I 12 10 Ten Year Fracture Risk...
Страница 412: ...This page intentionally left blank Operator s Guide 437D140 Rev I 14 16...
Страница 428: ...This page intentionally left blank Operator s Guide 437D140 Rev I 16 10...
Страница 446: ...Label Locations continued Operator s Guide 437D140 Rev I 17 18 Technical Reference...
Страница 448: ...Scanner Labels continued Operator s Guide 437D140 Rev I 17 20 Technical Reference...
Страница 449: ...Operator s Guide 437D140 Rev I Technical Reference 17 21 Scanner Labels continued...
Страница 457: ...Operator s Guide 437D140 Rev I Reference Data Sets 18 7 Excerpt from Calcified Tissue International...
Страница 458: ...Excerpt from the Journal of Bone and Mineral Research Operator s Guide 437D140 Rev I 18 8 Reference Data Sets...
Страница 459: ...Operator s Guide 437D140 Rev I Reference Data Sets 18 9...
Страница 466: ...Operator s Guide 437D140 Rev I This page intentionally left blank...
Страница 484: ...Operator s Guide 437D140 Rev I This page intentionally left blank...