![Norland Elite 437A150 Скачать руководство пользователя страница 210](http://html1.mh-extra.com/html/norland/elite-437a150/elite-437a150_operators-manual_1705413210.webp)
Maintaining High Quality Hip Scans
Patient positioning, scan and analysis techniques can influence the precision and accuracy of Norland Bone
Density estimations. Facilities can reduce the adverse effects of some of these factors by:
Performing and monitoring the daily QA procedure to verify that other radiation sources (x-ray machines, nuc-
lear imagers) are not affecting the performance of the Norland system. The daily QA procedure verifies
proper operation as well.
Ensuring that all operators position patients and analyze data in the same manner.
Screening patients for recent radionuclide uptake procedures. Residual emission may be misinterpreted by
the Norland system as x-rays.
Screening patients for recent ingestion of radiopaque substances. Barium or other dyes used in some x-ray
procedures could result in increased soft tissue x-ray absorption.
Screening patients for external opaque (metal and plastic) objects. Jewelry, buttons, zippers, rivets,
buckles, pens, keys etc. can affect the results if they are in the scanning region.
Screening patients for prosthetic devices, implants, buttock augmentation, surgical staples, or other high
density sub-dermal materials that may affect density estimates.
Ensuring that scan and analysis parameters remain constant for all scans of the same patient.
Always allowing the Hip Measure Scan to complete without interaction.
Ensuring the patient does not move during the scan.
Figure 8-2:
Example of a Good Quality Hip Scan
The femoral neck cursor should be a 1.5-cm horizontal slice of femoral neck, centered on axis, and should
not include any other anatomy.
The scan includes the entire Greater Trochanter, shows the Lesser Trochanter and some of the femoral
shaft.
The Lesser Trochanter is minimized in the region of interest, indicating proper rotation.
The Femoral shaft is straight, indicating proper abduction.
Operator's Guide
437D140 Rev. I
8-4
Scanning Hip
Содержание Elite 437A150
Страница 1: ...DXA Bone Densitometer Operator s Guide Elite Model 437A150 Operator s Guide 437D140 Rev I...
Страница 22: ...This page intentionally left blank Operator s Guide 437D140 Rev I 1 8...
Страница 41: ...Operator s Guide 437D140 Rev I General Information 2 19 Figure 2 6 Page 2 of the Bone Exam Report...
Страница 45: ...Operator s Guide 437D140 Rev I General Information 2 23 Figure 2 10 Page 2 of the Patient Letter...
Страница 47: ...Operator s Guide 437D140 Rev I General Information 2 25 Figure 2 12 Page 2 of the Referral Letter...
Страница 49: ...Operator s Guide 437D140 Rev I General Information 2 27 Figure 2 14 Page 2 of the Ten Year Fracture Risk Report...
Страница 56: ...This page intentionally left blank Operator s Guide 437D140 Rev I 2 34...
Страница 66: ...11 Click Operator s Guide 437D140 Rev I 3 10 Installation and Setup...
Страница 108: ...This page intentionally left blank Operator s Guide 437D140 Rev I 3 52...
Страница 124: ...Sample Printed QA Results Report Operator s Guide 437D140 Rev I 4 16 Basic Operation...
Страница 137: ...Operator s Guide 437D140 Rev I Basic Operation 4 29...
Страница 165: ...Operator s Guide 437D140 Rev I Scanning Whole Body 5 23 Figure 5 6 Page 2 of the Bone Exam Report...
Страница 166: ...This page intentionally left blank Operator s Guide 437D140 Rev I 5 24...
Страница 176: ...Figure 6 2 Page 2 of the Bone Exam Report Operator s Guide 437D140 Rev I 6 10 Soft Tissue Composition...
Страница 178: ...Figure 6 4 Page 2 of the Body Composition Report Operator s Guide 437D140 Rev I 6 12 Soft Tissue Composition...
Страница 198: ...Operator s Guide 437D140 Rev I 7 20 Scanning AP Spine...
Страница 205: ...Operator s Guide 437D140 Rev I Scanning AP Spine 7 27 Figure 7 5 Page 2 of the Bone Exam Report...
Страница 206: ...This page intentionally left blank Operator s Guide 437D140 Rev I 7 28...
Страница 238: ...Figure 8 6 Page 2 of the Bone Exam Report Operator s Guide 437D140 Rev I 8 32 Scanning Hip...
Страница 289: ...Operator s Guide 437D140 Rev I Scanning Lateral Spine 10 29 Figure 10 6 Page 2 of the Bone Exam Report...
Страница 290: ...This page intentionally left blank Operator s Guide 437D140 Rev I 10 30...
Страница 301: ...Operator s Guide 437D140 Rev I Additional Techniques 11 11 6 The dialog box closes and the file is sent...
Страница 326: ...Operator s Guide 437D140 Rev I 11 36 Additional Techniques...
Страница 378: ...Figure 12 2 Page 1 of the Ten Year Fracture Risk Report Operator s Guide 437D140 Rev I 12 10 Ten Year Fracture Risk...
Страница 412: ...This page intentionally left blank Operator s Guide 437D140 Rev I 14 16...
Страница 428: ...This page intentionally left blank Operator s Guide 437D140 Rev I 16 10...
Страница 446: ...Label Locations continued Operator s Guide 437D140 Rev I 17 18 Technical Reference...
Страница 448: ...Scanner Labels continued Operator s Guide 437D140 Rev I 17 20 Technical Reference...
Страница 449: ...Operator s Guide 437D140 Rev I Technical Reference 17 21 Scanner Labels continued...
Страница 457: ...Operator s Guide 437D140 Rev I Reference Data Sets 18 7 Excerpt from Calcified Tissue International...
Страница 458: ...Excerpt from the Journal of Bone and Mineral Research Operator s Guide 437D140 Rev I 18 8 Reference Data Sets...
Страница 459: ...Operator s Guide 437D140 Rev I Reference Data Sets 18 9...
Страница 466: ...Operator s Guide 437D140 Rev I This page intentionally left blank...
Страница 484: ...Operator s Guide 437D140 Rev I This page intentionally left blank...