M2000 Indicator Parameter List
The following gives a partial list of the M2000 parameters. The list explains
the use of the most commonly needed parameters on the indicator. However,
the M2000 features over 80 parameters. For a complete list contact Norac.
8.1 CAL INITIALIZATION Parameters
1 Reload Factory Default Values
This function will reinitialize the indicators scale calibration parameters for the
specific channel to the default factory values. When you have selected his
function you must confirm by pressing [1] followed by the [TARE] key. This
will only initialize a single channel. The indicator will automatically exit
calibration and reboot to initializing the calibration data to factory default. Any
other values will display error 7. The indicator will reboot after performing a
factory reset. Use parameter 260 below to reinitialize completely to factory
defaults.
260
Complete factory initialize
Similar to parameter 1 above. This parameter completely initializes all
indicator settings to factory defaults on all 3 channels. This includes COM port
settings and all other system related settings. Ticket data and tares stored in
memory are not initialized.
98 Enables scale channel (0)
This parameter enables scale channels 2 or 3. Channel 1 is always enabled and
cannot be disabled. By default channels 2 and 3 are turned off. To enable a
channel enter 1followed by the [TARE] key. To disable a channel enter 0
followed by the [TARE] key.
99 Exit calibration mode
This parameter will exit calibration mode and restart the indicator. All
calibration changes are stored to flash memory and the audit trail is updated.
8.2 SCALE SETUP Parameters
2 Decimal Point Position (0)
Sets the decimal position for the display. Values 0 to 4 can be entered
followed by the [TARE] key. An invalid entry will display error 3.
3 Graduation size (1d)
Selects the grad size to be used on the scale. Values that can be enter are 1, 2 ,
5, 10, 20, 50 and 100 followed by the [TARE] key. An invalid graduation size
will display error 2.
4 Scale Capacity (10000d)
This should be set to the scale capacity for the scale. For example if your scale
capacity is1000 lbs then you would enter would be 1000 followed by the
[TARE] key. The zero range window is calculated from this parameter