audit trail. The audit trail is permanent and cannot be disabled or erased by
removing the internal battery inside the indicator. The Audit Trail has two
counters, the Calibration counter and the Parameter counter. Changing
parameters that affect the calibration of weight will increment the
Calibration counter by 1. All other parameters will increment the Parameter
counter. The counters will count from 000 to 999 before rolling over again.
Changing one or more parameters will only increment the counter by 1 per
calibration session. Only an actual change of value to the calibration
parameter will register to the audit trail.
Important Note:
Because the audit trail becomes active at the factory, the counters may not
show 0 even when the indicator is new out of the box.
5.3.3
Accessing the Audit Trail
The audit trail can only be accessed from normal weighing mode and cannot
be accessed while you are in calibration mode. Enter 1000 followed by the
PRINT SELECT. The message “Audit” will briefly be displayed on the
display. Shortly after the indicator will cycles through 3 times displaying the
date of when the last change was made to the indicators calibration
parameters, the calibration (CAL) counter and the configuration (CFG)
counter. You may press the CLEAR key at any time to cancel this operation.