33
Technical Information
Equipment Response Time
If the signal from the sensor is inadequate, the last measured SpO
2
and pulse rate values freeze
for 10 seconds and are then replaced with dashes.
SpO
2
Values
Average
Latency
Standard/Fast Averaged SpO
2
4 beat exponential 2 beats
Pulse Rate Values
Response
Latency
Standard/Fast Averaged Pulse Rate
4 beat exponential 2 beats
Equipment Delays
Delay
Display Update Delay
1.5 seconds
Alarm Signal Generation Delay
0 seconds