Table 7. Electromagnetic Immunity
Immunity Test
Compliance Level
This device is intended for use in the electromagnetic environment specified in the Indications For Use.
Electrostatic Discharge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
Electrical Fast Transient/Burst
IEC 61000-4-4
±2 kV for power supply lines
±2 kV for input/output lines
Surge
IEC 61000-4-5
±0.5kV, ±1kV, ±2kV for Line-to-Ground
±0.5kV, ±1kV for Line-to-Line
Voltage dips, short interruptions, and
voltage variations on power supply
input lines
IEC 61000-4-11
0% UT for 0.5 Cycles at 50 Hz at 0, 45, 90, 135, 180, 225,
and 315° Phase
0% UT for 1 Cycle at 50 at 0°Phase
70% UT for 25 Cycles at 50 Hz at 0°Phase
0% UT for 250 cycles at 50 Hz at 0°Phase
Power Frequency (50/60 Hz) Magnetic
Field
IEC 61000-4-8
30 A/m
Conducted RF IEC 61000-4-6
150 kHz to 80 MHz
3 Vrms
ISM and Amateur radio bands
between 150 kHz to 80 MHz
6 Vrms
Radiated RF IEC 61000-4-3
80 MHz – 2.7 GHz
10 V/m
380 – 390 MHz
27 V/m
430 – 470 MHz
28 V/m
704 – 787 MHz
9 V/m
800 – 960 MHz
28 V/m
1.7 – 1.99 GHz
28 V/m
2.4 – 2.57 GHz
28 V/m
5.1 – 5.8 GHz
9 V/m
Note:
U
T
is the AC mains voltage before application of the test level.