17
Operator Functions
NOTE:
Alarm limits cannot be changed when the Model 7500FO is in Patient Security mode.
Patient Security mode prevents accidental changes to critical parameters. The Model 7500FO
allows users to lock and unlock alarm limits, volume settings, and time settings.
Enter Patient
Security Mode
To enter Patient Security mode, press and hold the
Alarm Silence button while turning on the device.
Exit Patient
Security Mode
To exit Patient Security mode, press and hold the
Alarm Silence and Limits buttons while turning on
the device.
Make Current
Alarm Values
User-defined
Defaults
To set the User-Defined Defaults to the current
alarm settings, hold the Alarm Silence button and
then press the Limits button.
Revert to Factory
Defaults
To revert to the factory defaults, from the User-
Defined Defaults alarm limits, hold the alarm
Silence button and then press the Minus (-) button.
NOTE:
User-defined default values will be lost
when factory defaults are made active.
CAUTION:
Review all limits to ensure they are appropriate for the patient.
Table 5: Advanced Functions (Continued)
Function
Button
Instruction
+
+
+
+
+
!