Reviewing and Setting Volume and Alarm Limits
Recalling Previous Settings
The pulse oximeter includes a feature that allows recall of the operator-adjusted settings in use
when the device was last turned off. The following settings are recalled when this feature is
activated:
• SpO
2
high and low alarm limits
• Pulse rate high and low alarm limits
• Alarm volume settings
Previous operator-adjusted settings can be recalled by pressing the Limits button while the unit is
on. “
rCL
” appears, indicating that previous alarm limit settings may be recalled. To recall the
settings, press the Plus button and select “
yES
.” Press the Limits button again to accept the recall
and return to normal operation.
NOTE:
The recalled value for the SpO
2
low alarm will not be less than the current default.
Error Codes
This device includes error codes that indicate problems with the unit. Error codes are indicated by
“
Err
” in the %SpO
2
display, and a capital “
E
” followed by a 2-digit code in the pulse rate display.
To correct error conditions, perform the following steps:
1. Turn the unit off and then back on again to remove the error code.
2. If the error persists, disconnect all power, and then reconnect the power and turn the unit
back on.
If the error still persists, note the error code and contact Nonin Technical Service at
(800) 356-8874 (USA and Canada), +1 (763) 553-9968, or +31 (0)13 - 79 99 040 (Europe).
CAUTION:
Review all limits to ensure they are appropriate for the patient.
!