mia505en2102p
10 / 16
3.1.
M
EASUREMENT DATA
(MSZ
EN
60770-1:2011,
MSZ
EN
61298-1:2009,
MSZ
EN
61298-3:2009)
T
YPE
M
–5
–
Resolution (on HART
®
transmitted value)
1 mm (0.04")
Nonlinearity (on HART
®
transmitted value)
±2 mm (±0.08") or ±0.085% whichever is greater
Under reference conditions
Hysteresis (under reference conditions)
±0.25 mm (±0.01")
Zero span (in LEVEL measurement mode)
Anywhere within the active range
Measurement range (reducing)*
Minimal range: 32 mm (1.26"),
Maximum range: see “3.2 Dimensions”
Temperature error
0.04 mm (0.0016") / 10 °C (18 °F) (between
−
25…+50 °C [
−
13…+122 °F])
Current output resolution
0.4
μ
A
Current output accuracy
33
μ
A
Current output temperature error
6 ppm / °C
* The accuracy data is only valid for factory default settings!