htk4014a600p_04
65/68
5.5.
T
ROUBLESHOOTING
Event
Fault
Action
Error messages
“Tank full” status marker on*,
reading frozen at max. or min.
value
No fault. The level has reached (and possibly risen
above) the top configured measurement limit and is
either displaying the maximum (when measuring
level) or minimum (when measuring distance) output.
None. Measurements should be normal once the level is in the configured
measurement range.
“Tank empty” status marker
on*, reading frozen at max. or
min. value
No fault. The level has entered the gauge’s bottom
dead zone and can no longer detect a return signal.
Either the maximum (when measuring distance) or
minimum (when measuring level) output is displayed.
None.
Measurements should be normal once the level is in the configured
measurement range.
“Tank full” and “Level lost”
status marker on*, reading
frozen at max. or min. value
No fault. The level has entered the gauge’s top dead
zone and can no longer detect a return signal.
Empty the tank below the top measurement range limit and
check the measurement.
“Level lost” status marker on*,
reading is frozen
The instrument has lost the level signal, has searched
but not yet found the return pulse. This may occur if
the pulse has dropped below the threshold. Parasite
signals from the flange or obstructions in the tank may
render the gauge unable to identify the correct signal.
Ensure that tank is emptied below maximum level and check the
measurement.
If the signal is not detected then modify the control threshold manually as
shown in sections 5.1.2.4.2 using the oscilloscope diagram and threshold
set functions (5.1.2.4).
“Reference not found” status
marker on*
Occurs when there is a problem with the time base on
the HF board.
Please contact NIVELCO.
“Level lost” and
“Reference not found” status
markers on*, reading frozen
The probe has received an electrostatic discharge.
The gauge will search for the level again and resume readings. If the
reading remains frozen then the signal converter may have been damaged
by ESD and may need replacing. Please contact NIVELCO.
“Flange not found” status
marker on*
The signal converter has been incorrectly configured
to measure with a cable or rod probe when it is
equipped with a coaxial probe. This may be also due
to installation on a long nozzle which has the effect
of attenuating the flange pulse.
Contact NIVELCO for the corrective procedure.