htk4014a600p_04
26/68
4.4.
P
OWER
-
ON AND START
-
UP
The MicroTREK 2-wire is pre-configured in accordance with the customer order specifications and ready for use immediately.
Attention! A start-up time of less than 60 seconds should be allowed once the unit is connected and the power is turned on.
If the probe length has been shortened, please refer to the following sections:
3.3.1: Summary of User Functions, user function 1.1.6: Probe length to modify configured probe length.
4.5.
A
VAILABLE USER INTERFACES
Programming of MicroTREK can be done using the following instruments / accessories:
SAP-300 display unit
Optional. See chapter „5.1. Programming with SAP-300 display unit”.
MultiCONT universal process controller
To be ordered separately. For programming instructions see the User Manual of MultiCONT.
HART
®
(HHC) Handheld Communicator
Separately available. Automatic device detection upon power-on.
See chapter „5.3 Programming with HART
®
(HHC) Communicator”.
5.
PROGRAMMING
MicroTREK can be programmed in three (basic) ways:
Programming
with EView2 software
Programming with SAP-300 display unit
Programming with HART
®
Handheld (HHC) Communicator
(For operating instruction see the User manual of the HHC)
5.1.
P
ROGRAMMING WITH
EV
IEW
2
SOFTWARE
5.1.1.
EView2 software installation and execution.
Install the “EView2 HART
®
configuration software” (hereinafter “EView2”) as described in chapter 3 of the EView2 software user’s manual.
Electrical connections: connect the transmitter to the PC using a HART
®
adapter (can be ordered separately).
In hazardous-duty systems a HART
communication capable Ex repeater must be inserted in the circuit before the interface!
The HART adapter should be connected to the intrinsically safe input of the Ex repeater!
Run the application and detect the devices with the software (see: Chapter 4 of the EView2 User's Manual for more details).
Содержание MicroTREK HB