Installation of two devices
Process connection and entry pipe
Stilling wells
If two devices to be used on the same
tank, these should be mounted at a
distance of at least 2 m (6.5 ft) from each
other. If this is ignored, interferences from
the electromagnetic (EM) fields generated
by both instruments may cause
measurement errors.
Coaxial probes: the outer shell of the
probe contains the EM field: no minimum
distance required.
Caution: Do not put the nozzle close to the entry
pipe.
Pouring the medium directly onto the probe will give
false readings.
Install deflector plate if impossible to distance gauge
from entry pipe.
Tanks with floating roofs for petrochemical applications:
Use a stilling well.