P12: - - b a Current output mode
FACTORY DEFAULT: 0
a
Error indication by output current
Error indication by output current:
Errors will be indicated by the
EasyTREK Pro
transmitter on the current output according to the set
value as long as error is present.
EView2 setting:
“Device Settings”
→
“Outputs”
→
“Current output”
→
“Error indication by the current output.”
0
HOLD (
hold last value)
1
3.8 mA
2
22 mA
b
Current output mode
Automatic:
The current loop output value is calculated from the measured value; the transmitter
output is active.
Manual:
The current loop output value is not calculated from the measured value, but a fixed
(according to P08) current loop output value is transmitted. In this mode, the current output error
setting is irrelevant.
This parameter overrides the HART multidrop communication mode 4 mA value!
0
Automatic
1
Manual
c
Initial output current
LoSUC - Lo StartUp Current:
The output current is calculated from the measured value; the
transmitter output is active. The output current is low (3.5 mA) at start-up.
HiSUC - Hi StartUp Current:
The output current value is calculated from the measured value; the
transmitter output is active. The output current at start-up is high (23.5 mA).
Caution! This setting is not compatible with HART multidrop mode!
0
LoSUC (3.5 mA)
1
HiSUC
(23.5 mA)