Start the program and search for the transmitter in the program (for more details, see EView2 user manual chapter 4).
Select a device from the list of detected devices and open the “Device programming” window of the device (chapters 4.4 and 4.5 of the EView2 user manual).
EView2 lets you set all the parameters and functions. The following chapter summarizes where and how to access the parameters.
5.3.2
M
EASUREMENT CONFIGURATION
P00: - c b a Engineering Units
FACTORY DEFAULT: 000
The device loads the entire parameter set with the factory parameter values according to the new unit system if the parameter is changed.
Therefore, all parameters must be set again!
a
Operation
The device is only suitable for measuring liquid levels, so this parameter cannot be changed.
Eview2 setting:
“Device Settings” → “Application” → “Operating mode
.”
0
Liquid level measurement
b
Engineering units
(according to “c”)
The unit of measure of the measured value when the instrument is set to measure distance or
level. It is also the unit of measure for distance or length type parameters.
EView2 setting:
“Device Settings” → “Application” → “Engineering uni
ts.”
Metric (EU)
Imperial (US)
0
m
ft
1
cm
Inch
c
Unit system
The used units of measurement (length, volume, mass) are determined by the unit system
set in this parameter. When changing the units of measure, the unit system must be selected
first, and only then can the actual unit be set.
EView2 setting:
“Device Settings”
→
“Application”
→
“Calculation system.”
0
Metric (EU)
1
Imperial (US)