Nitto Construction
- 36 -
STR
Strength index value with each
impact
INDX
Surface deterioration index value
with each impact
STAT
Delamination index value with
each impact
・
When the first hammering is completed, it is not determined whether the data is adoptable
or not. MN and N remain 0. If there is nothing wrong with the data, hammering should be
continued.
・
At the second hammering, the data of the first hammering is recorded and the mean score
and the numbers of measurement are renewed.
・
If there is something wrong with the measured data, press
button twice so that the
measured data becomes ineffective and is deleted from measured waveform.
(
When
pressing
button and followed by pressing
button, judgement of ineffectiveness is
released.
)
・
When finishing the measurement, press
button. (At this stage, the number of the
measured point and the mean value are settled.)
※
Please note that the data of the last waveform is not reflected in calculation of the mean
score unless
button is pressed. When pressing
button, the screen will return to
LOOKUP mode that the recorded data exists (See Figure 7.2-2).
・
The data cannot be recorded if the impact was too strong or too light. in CTS-02V4. Please
check the impact strength by trying to hammer a few times prior to a real hammering.
After hammering, electronic sound indicates if it is a proper hammering or not
*Bleep once: Normal, Bleep three times: Too strong, No bleep: Too light
7.4 Deletion of Data
・
CTS02V4 body can delete the data with each coordinate number.
・
When selecting the coordinate number that you want to delete at LOOKUP mode (See
Fig.7.2-2) and pressing
button, it sounds the beep and the screen of display changes as
shown in Fig.7.4-1.
・
In this state, all data recorded in the specified coordinate number will be deleted by pressing
button.
・
When you don’t want to delete, press
button and return to LOOKUP mode (See
Fig.7.2-2).
*You cannot delete all data in SITE at once.
Содержание CTS-02V4
Страница 1: ...CTS 02V4 User Manual...