33
Messages
Display
Reason
Meter
Error
Temperature
Error
•
Too Cold/
Too Hot
Sample Not
Detected or
Using Wrong
Test Strip
Used Test
Strip,
Test Strip
Outside of
Vial too Long
Meter
Error
Test Strip Error, Very
High Blood Glucose
Result (higher than 33.3
mmol/L (600 mg/dL).
Action
Call for
assistance.
Move Meter and Test Strips
to area between 10°C-40ºC;
wait 10 minutes for System
to reach room temperature
before testing.
Retest with new
Test Strip and
larger sample.
Repeat with new Test Strip.
Make sure sample is touched
to edge of Test Strip (not
top). If error persists,
call for assistance.
Call for
assistance.
Retest with new Test Strip. If error persists,
call for assistance. If you have symptoms such
as fatigue, excess urination, thirst, or blurry
vision follow your healthcare professionals
advice for high blood glucose.