29
CARE, CLEANING
Caring for TRUEresult twist 2
• Store System (Meter, Control Solution, Test
Strips) in Carrying Case to protect from liquids,
dust and dirt. Do not keep meter in an area
where it may be crushed (i.e. back pocket,
drawer, bottom of bag, etc.).
• Store in a dry place at 2°C-30°C
(room temperature).
•
DO NOT REFRIGERATE
OR FREEZE.
Meter Care
Cleaning removes blood and soil.
To Clean the Meter
:
• Clean the Meter when visibly dirty or if blood is
on the Meter.
• Never put Meter in liquids or allow any liquids to
enter the Test Port.
• Wipe Meter with a clean, lint-free cloth
dampened with 70% Isopropyl alcohol.
• Let Meter air dry thoroughly before using to test.
• Do not use bleach to clean the Meter.
For assistance call the phone number on the cover of
the Owner’s Booklet.
2°C
30°C