14
GETTING STARTED
The Meter turns on when a Test Strip is inserted
into the Test Port or when Set Button is pressed and
released (see
Memory
). Meter turns off when the
Test Strip is released from the Meter, Set Button is
pressed and held for 20 seconds, or after 2 minutes
of non-use.
Always check your supplies before using.
• Check Meter for damage (cracked Display,
missing buttons, etc.). If damage is seen, do not
use Meter. Call for assistance.
• Check Test Strip vial for damage (cracked vial,
broken vial, etc.). Discard damaged vial and all
contents (Test Strips) and use a new vial of Test
Strips for testing.
• Write date first opened on Test Strip vial. Check
Use By Dates (written and printed) before using
any Test Strips from the vial. Do not use if 4
months after opening (written date) or if printed
Use By Date has passed.
• Check Control Solution bottle for any leaks or
broken cap. Discard bottle and open a new one
for testing.
• Write date first opened on Control Solution
bottle label. Check Use By Dates (written and
printed) before using. Do not use if 3 months
after opening (written date) or if printed Use By
Date has passed.