SYSTEM SPECIFICATIONS
Result Range:
1.1 - 33.3 mmol/L (20-600 mg/dL)
Sample Size:
Minimum 0.5 microlitre (0.5 μL)
Sample:
Fresh capillary whole blood, venous blood
drawn in sodium or lithium heparin blood
collection tubes, or Control Solution.
Test Time:
Results in as little as 4 seconds
Result Value:
Plasma equivalent values
Assay Method:
Electrochemical
Power Supply:
One 3V Lithium Battery
#CR2032
(non-rechargeable)
Battery Life:
Approximately 1,500 tests or 2 years
Automatic shut-off:
After two minutes of non-use
Weight:
17 grams
Size:
4.3 cm x 3.7 cm 2.3 cm
Memory Size:
500 glucose results
Operating Range
(Meter & Test Strips):
Relative
Humidity:
10-90% (Non-condensing)
Temperature:
10°C-40°C
Haematocrit:
20-55%
Altitude:
3094 metres
Note:
Use within specified
environmental conditions only.
Chemical Composition
TRUEresult Test Strips:
Glucose dehydrogenase-FAD
(Aspergillus sp.)
, Mediators, Buffers and Stabilisers.
TRUEresult Control Solution:
Water, d-glucose,
buffers, viscosity enhancing agents, salts, dyes and
preservatives.
39