Chapter 4 Assembly
61
Installing the eyepiece tube
Insert the eyepiece tube into the circular dovetail
mount while tilting it so that the eyepiece tube fits into
it properly.
Securely tighten the fixing screw for the eyepiece tube
while supporting the eyepiece tube by hand.
Fit the auxiliary fixing screw for the eyepiece tube and
tighten it by using the supplied hexagonal wrench.
Function of the auxiliary fixing screw for
the eyepiece tube
The auxiliary fixing screw for the eyepiece tube is for
preventing the eyepiece tube from falling when the
fixing screw for the eyepiece tube is loosened. The
eyepiece tube cannot be fastened only with the
auxiliary fixing screw for the eyepiece tube.
When screws are attached
If you purchase the microscope without the eyepiece
tube, the fixing screw and auxiliary fixing screw for
the eyepiece tube are fitted into the screw holes and
taped to prevent them from falling. When installing
the eyepiece tube, peel off the tape and remove the
two attached screws, and then follow the procedure
described in “
■
Installing the eyepiece tube”.
3
Attach the camera mount
The camera mount (C mount) is attached to the
camera port of the trinocular eyepiece tube.
Loosen the knurled screw on the camera port and
remove the cap and C mount. Photomicrography
becomes available after attaching the C mount to the
camera, and then attaching it to the camera port. The
optical path of the trinocular eyepiece tube is fixed at
the ratio: binoculars : camera port = 50 : 50.
For details about photomicrography, refer to the
instruction manual of the camera.
Do not lose the cap. When you do not attach the
camera, make sure that the cap is placed on the
camera port.
Camera mount
(C mount)
Knurled screw
The cap is attached.