Chapter 7 Specifications
69
Chapter 7
Specifications
7
Specifications
7.1 Microscopy
(Principles)
Use the objective and eyepieces of the microscope to magnify minute cells and tissue optically, manipulate
the lever and knobs of the microscope unit to adjust the focus or move the observation point. Then, observe
the sample affixed to the slide.
Intended use of the product (for medical care)
This microscope is intended for use in microscopic experiment and diagnostics of cells and tissues at
hospitals or by doctors in private practice in the field of pathology, anatomy, and cytology.
The microscopy with diascopic illumination is used to observe a sample fixed on the slide (cells and tissue) as
the specimen.
The product is classified as an in-vitro diagnostic medical device.
IVD
This product is not intended for use for measurement.
The scale on the focus knob and stage is an indicator to reproduce the position and does not guarantee the
value of the thickness or length of a sample measured using this scale.
Intended user
It is intended for the medical professional and those who work on experimentations in the field of pathology
and cytology.
7.2 Performance
Properties
Model Name:
Main Body: ECLIPSE Ei R
Optical System:
CFI45 (Infinity-corrected CF optical system)
Second objective focal length f = 200 mm
Built-in diascopic illumination system (Simplified Kohler’s illumination system)
Focusing Mechanism:
Fine focus knob graduation:
2
μ
m/graduation
Fine focus knob travel:
0.2 mm up or down / revolution
Coarse focus knob travel:
37.7 mm up or down / revolution
Stage vertical movable range: From the focal plane
2 mm upward and 13 mm downward
Stage:
Stroke: Right and left axis:
76 mm
Front and back axis:
30 mm
Revolving Nosepiece:
4-hole fixed type
Illumination Light Source:
LED