17
RS-232C Communication and USB Communication
:
Data Format
2
MLM26*RTZ002F
When the Prism parameter is set to P-B:
The measurement data is output as fixed digits with the leading digit 0 remaining.
1) Example of single data when the Prism parameter is set to “BU/D BI/O”:
(PRISM = IN 3.00
Δ
, UP 2.50
Δ
)
2) Example of right data when the Prism parameter is set to “BU/O BI/O”:
(PRISM = OUT 1.25
Δ
, DOWN 2.00
Δ
)
3) Example of left data when the Prism parameter is set to “P-B”: (PRISM = 6.5
Δ
, BASE = 70°)
(9)Progressive length data (LM-1800PD only)
The measurement data of progressive length (V. Length) is output.
The measurement data of progressive length is output as two-digits with the leading digit 0
remaining.
1) Example of single data (V. Length= 12 mm)
Identification
code
Single
“P ” “B ” (P, Space) (B, Space)
Right
“PR” “BR”
Left
“PL” “BL”
PRISM
Four digits with a fixed decimal point
with no polarity sign
BASE
Three-digit integer with no polarity sign
P
0
3
.
0
0
I
ETB CR
P
0
2
.
5
0
U ETB CR
50H 20H 30H 33H 2EH 30H 30H 49H 17H 0DH 50H 20H 30H 32H 2EH 35H 30H 55H 17H 0DH
PRISM1 (6 characters)
PRISM2 (6 characters)
BASE IN
BASE UP
P
R
0
1
.
2
5
O ETB CR
P
R
0
2
.
0
0
D ETB CR
50H 52H 30H 31H 2EH 32H 35H 4FH 17H 0DH 50H 52H 30H 32H 2EH 30H 30H 44H 17H 0DH
PRISM1 (6 characters)
PRISM2 (6 characters)
BASE OUT
BASE DOWN
P
L
0
6
.
5
0
ETB CR
B
L
0
7
0
ETB CR
50H 4CH 30H 36H 2EH 35H 30H 17H 0DH 42H 4CH 30H 37H 30H 17H 0DH
PRISM (5 characters)
BASE
(3 characters)
Identification
code
Single
“D ” (D, Space)
Right
“DR”
Left
“DL”
Содержание LM-1800P
Страница 1: ...INTERFACE MANUAL MLM26 RTZ002F LM 1800P LM 1800PD AUTO LENSMETER...
Страница 6: ...2 OUTLINE OF COMMUNICATION Outline of Communication Function MLM26 RTZ002F...
Страница 54: ...50 RS 232C Communication and USB Communication Example of Data Output MLM26 RTZ002F...
Страница 70: ...66 COMMUNICATION PROCEDURE other than NCP20 Flowchart MLM26 RTZ002F...