19
RS-232C Communication and USB Communication
:
Data Format
2
MLM26*RTZ002F
2) Example of the right data (Green = 75%)
(12)UV transmittance data
The measurement data of UV transmittance is output.
The measurement data of UV transmittance is output as three-digits with the leading digit 0
remaining.
1) Example of the single data (UV = 20%)
2) Example of left data (UV = 15%)
G
0
8
0
ETB CR
47H 20H 30H 38H 30H 17H 0DH
Green
(3 characters)
G
R
0
7
5
ETB CR
47H 52H 30H 37H 35H 17H 0DH
Green
(3 characters)
Identification
code
Single
“U ” (U, Space)
Right
“UR”
Left
“UL”
U
0
2
0
ETB CR
55H 20H 30H 32H 30H 17H 0DH
UV
(3 characters)
U
L
0
1
5
ETB CR
55H 4CH 30H 31H 35H 17H 0DH
UV
(3 characters)