12
RS-232C Communication and USB Communication
:
Data Format
MLM26*RTZ002F
The LM transmits the following types of data. These types of data are output according to the lens
measurement mode, whether the data has been measured or not, and the condition for the lenses
(single, left or right).
: Output if measured ×: Not output
*1: LM-1800PD only
Only when the parameter is set to “On”, the measurement is performed and data is output.
*2: 1st add power only
*3: Add power in negative value is sent as positive value.
Ex) SPH -2.00 ADD -1.00
→
SPH -3.00 ADD 1.00
*4: In the case of “single (not the right- nor left-eye lens)” state, only single data is output.
When both left- and right-eye lenses are measured, the data is output in following order:
“Right” “Left”.
In the order of “1) - 4)”
→
“Right 5) - 15) data”
→
“Left 5) - 15) data”
Measured lens mode
Prism
layout
Lens
*4
Single
vision
Progressive
power
Contact
lens
Single Right
Left
1) Company name/Model name
Once at the beginning
2) Operator/ Patient ID
Once at the beginning
3) Patient number (print number)
Once at the beginning
4) Date and time
Once at the beginning
5) SPH, CYL, AXIS
6) SE (Spherical Equivalent value)
×
×
×
7) ADD (1st/2nd add power)
*3
*2
×
×
8) PRISM
9) Progressive length
*1
×
×
×
10) Channel width
*1
×
×
×
11) Green transmittance
12) UV transmittance
13) PD
×
×
14) Inside amount
*1
×
×
×
×
15) Net prism
*1
×
×
×
×
• The measurement data has been rounded off according to the cylinder mode, the indication
form of the measured prism data, and indication step specified on the LM.
Содержание LM-1800P
Страница 1: ...INTERFACE MANUAL MLM26 RTZ002F LM 1800P LM 1800PD AUTO LENSMETER...
Страница 6: ...2 OUTLINE OF COMMUNICATION Outline of Communication Function MLM26 RTZ002F...
Страница 54: ...50 RS 232C Communication and USB Communication Example of Data Output MLM26 RTZ002F...
Страница 70: ...66 COMMUNICATION PROCEDURE other than NCP20 Flowchart MLM26 RTZ002F...