6. Maintenance
RH-PE
–
Mercury collection-tube reheating furnace for the PE-1 19
No.600-2113-03
6.1.3.
Abnormal blank value
The blank value is abnormally high (e.g. several ng) and barely decreases.
Carry out maintenance of the PE-1.
Wash or replace the V5, joint, Teflon
tubes, or H5 collection tube.
Wash the gas washing bottle,
dehumidifying bottle, joint, and
coupling.
START
END
NO
Does the blank
value decrease?
YES
NO
Replace the transparent
Teflon tube.
Does the blank
value decrease?
NO
YES
Does the blank
value decrease?
YES
NO
Replace the septum.