5. Measurement
14
Mercury/RH-PE
‒
Mercury collection-tube reheating furnace for the PE-1
No.600-2113-03
5.4.4.
Measuring the content of mercury collection tube
(1) Make the setting for an unknown sample in the Table window. Check the “Meas” field,
and select “23:RH” in the “Method Number” field.
(2) Insert the mercury collection tube to be measured into the coupling from the top of the
RH-PE in such a way that its narrow section is pointed downward (is on the lower end).
(3) Connect the joint with the septum to the upper end of the mercury collection tube
(4) To commence measurement, select [Run]-[Start] from the menu, or click the
icon.
Alternatively, press F5 on the keyboard to start measurement.
(5) The following message appears when measurement is complete.
Remove the joint with the septum from the measured mercury collection tube, and pull
it out of the RH-PE.
(6) To continue measurement, return to step (1) and repeat the procedure as necessary.
Gas flow
Narrow
section