5. Measurement
16
Mercury/RH-PE
‒
Mercury collection-tube reheating furnace for the PE-1
No.600-2113-03
5.4.5.
Measurement sequence
To commence measurement, select [Run]-[Start] from the menu, or click the
icon.
Alternatively, press F5 on the keyboard to start measurement.
The measurement sequence proceeds as follows:
Operation Status
name
Period
1
Heating of the
collection tube
SAMPLE 120
seconds
2
Ar substitution 1
PURGE1
30 seconds
3
Ar substitution 2
PURGE2
30 seconds
4
Ar substitution 3
PURGE3
10 seconds
5
Ar substitution 4
PURGE4
30 seconds
6
Measurement MEASURE
Period
specified in the measurement parameter
7
Cooling COOL
30
seconds
5.4.6.
Measurement interruption
To interrupt the measurement, select [Run]-[Stop] from the menu or click the
icon.
Clicking the icon terminates operation on completion of the current measurement.
Clicking the
icon immediately starts the cooling process, and the equipment
returns to READY status when cooling has completed.
6. Maintenance
6.1.
Criteria for judging equipment failure and troubleshooting
Listed below are criteria for judging whether the equipment is operating properly or not.
(If the equipment does not satisfy any of these criteria, take the appropriate measures
given in "Troubleshooting" described in Sections 6.2.1 to 6.2.3.)
(1) A blank value should be low (0.01 ng or less).
→
6.1.3
(2) After a standard mercury gas is measured, the blank value should not significantly
increase. (As a guide it should be approximately 1% or less of the mercury measured
right before.)
→
6.1.3
(3) When a standard mercury gas is repeatedly measured its reproducibility should be
within
±
3%.
→
6.1.2
(4) A calibration curve drawn using a standard mercury gas should be linear with a
variation of
±
5% or less.
→
6.1.2
(5) When a calibration curve is drawn for a low mercury weight region (1 ng or less), the
y-intercept of the curve should not be negative.
→
6.1.1
(6) When a standard sample is measured before and after measurement for the collection
tube, the difference between the measured values should be within the measurement
error.
→
6.1.1
(7) The intensity of fluorescence should be at least 70% of that at the time of delivery.
→
6.1.1