Chapter 4
Calibration
©
National Instruments Corporation
4-3
DAQPad-MIO-16XE-50 User Manual
perform an external calibration very often. You can externally calibrate the
device by calling the NI-DAQ
Calibrate_E_Series
function.
To externally calibrate the device, be sure to use a very accurate external
reference. The reference should be several times more accurate than the
device itself. For example, to calibrate a 16-bit device, the external
reference should be at least ±0.001% (±10 ppm) accurate.
To manually calibrate the DAQPad-MIO-16XE-50, refer to the
E Series
Calibration Procedure
at
ni.com/calibration
.