Hunan Next Generation Instrumental T&C Tech. Co., Ltd.
NGI
l
NGI
N36100 Series
Wide Range High Voltage Programmable DC Power Supply
42
Table 25
Fault
Possible Causes
Solutions
There is no output after
switching on.
The
voltage/current
setting is zero.
Set
the
required
voltage/current.
The voltage output stops
after transient output.
There is over voltage
protection (OVP).
Lower the setting.
It is not allowed to set output voltage/current properly.
Table 26
Fault
Possible Causes
Solutions
The
output
voltage/current cannot be
set properly.
The set voltage/current is
higher
than
V-Limit
Max/I-Limit Max.
Set
the
required
voltage/current.
The set voltage/current is
lower
than
V-Limit
Min/I-Limit Min.
Unstable output voltage.
Table 27
Fault
Possible Causes
Solutions
The output voltage are not
stable.
N36100
is
switching
between CV and CC.
Change the setting.
The output voltage is
wobbly.
The sense terminal is not
connected.
Connect
to
sense
terminal, using the green
terminal supplied.
High output voltage ripple.
Table 28
Fault
Possible Causes
Solutions
The
ripple
is
high
sometimes.
The output voltage is out
of range.
Adjust the output voltage.
The ripple is high in
another
operation
environment.
There
is
strong
electromagnetic
interference nearby.
Stay away from the
interference.