N36100 Series
Wide Range High Voltage Programmable DC Power Supply
28
Table 17
OVP
This parameter is used to set over voltage protection value. Once the output voltage
exceeds the OVP set value, N36100 will immediately shut off the output and protect
the DUT. Meanwhile, alarm OVP will be displayed on the screen.
OCP
This parameter is used to set over current protection value. Once the output current
exceeds the OCP set value, N36100 will immediately shut off the output and protect
the DUT. Meanwhile, alarm OCP will be displayed on the screen.
OPP
This parameter is used to set over power protection value. Once the output power
exceeds the OPP set value, N36100 will immediately shut off the output and protect
the DUT. Meanwhile, alarm OPP will be displayed on the screen.
Steps to set parameters under
Protection
:
Rotate
to select the required parameter→ Press
on the required
parameter→ Press
to move the cursor→ Rotate
to adjust the parameter →
Press
Enter
button to confirm the setting.
7.8 System
Methods to enter
System
:
Method 1: Press
Shift
button first on the front panel and then press
Menu
button.
Method 2: Press
Menu
button→ Select
System
by rotating
→ Press
Enter
button.
Parameter
Function
OVP
To set over voltage protection value
OCP
To set over current protection value
OPP
To set over power protection value
Note: To exit from
APG
, please press
Menu
button or the function button required.