N36100 Series
Wide Range High Voltage Programmable DC Power Supply
23
Table 13
7.3.1 Parameter Setting
Steps to set parameters under
SEQ Edit
:
Rotate
to select the required parameter→ Press
on the required
parameter→ Press
to move the cursor→ Rotate
to adjust the parameter →
Press
Enter
button to confirm the setting.
7.4 SEQ Test
SEQ test (sequence test function) supports simulation of complex voltage & current
waveform, which is frequently used for automotive electronics test, engine start-up
test, etc.
The principle of sequence test is to output the voltage and current according to the
test steps edited by the user. When dwell time reaches, it will switch to the next step.
Parameter
Function
File
To set the sequence file number, Max. 10 files
Steps
To set the total test steps for the SEQ file, Max. 200 steps
Cycles
To set the number of cycles for the file under edit
Link to File
Link to the required file after the present file is
completed. Zero means no link.
Step No.
To set the test step number, Max. 200 steps
V-Set
To set the output voltage for the present step
I-Set
To set the output current for the present step
Dwell
To set single step delay time for the present step
Note: To exit from
SEQ Edit
, please press
Menu
button or the function button