PSM3750 Quick User Guide
Page 64
Single Compensation:
Selecting single compensation will
conduct zero compensation, short or
open circuit at the same desired
frequency that the test will be carried
out at.
Sweep Compensation:
Sweep compensation needs to be
configured exactly the same as the
sweep details for the DUT to be
tested. Frequency start, frequency
end and steps all need to be
replicated.
If “auto shunt” is selected
within the LCR menu which it will be
by default then sweep compensation
will be carried out on each individual
shunt in turn over the sweep
parameters
set
earlier
Note: When undertaking open compensation, it is advisable to use a solid piece
of wire to connect to each Kelvin Lead or on each side of the Kelvin Fixture.
For short circuit compensation, it is advisable to use a solid piece of wire to
connect the two Kelvin Leads together or connect between the two sides of the
Kelvin Fixture.
Note: If any compensation has been undertaken and the resultant program
saved along with all the measurement settings then upon recall of this program
the same compensation settings will still be applicable.