PSM3750 Quick User Guide
Page 56
If the PSM3750 displays the following message:
Then check that the extension port cable has not been damaged and has been
correctly fitted between the IAI2 and the PSM3750.
I
f the message does not appear but all 4 led’s remain illuminated and flashing
then check that the IAI2 is correctly connected to the supply, switched on at the
rear and the fuse is intact.
To test that the IAI2 is responding, connect across the Kelvin leads an
appropriate test component, for example a 22
0Ω resistor
and switch on the
PSM’s generator from the “OUT” menu. Press the “LCR” key and make sure that
the test component is measured correctly as shown below.