PSM3750 Quick User Guide
Page 45
8. PSM3750
–
Measurement Functions:
8.1 FRA - Frequency Response Analyzer
PSM3750 measures the gain and phase of channel 2/3 (dependant on model)
relative to channel 1 using a Discrete Fourier Transform (DFT) algorithm at the
fundamental frequency.
Test Equipment: 1 x SMPS (Switch Mode Power Supply) Test Box
1 x PSM3750
Enter “FRA” mode by pressing the
FRA button. Use the arrow until
the red box surrounds the gain/phase
margins parameter now press
arrow to enable this function, these
parameters will now be displayed
within the graph results display as
shown later in this section.
Enter “SWEEP” control mode by
pressing the Sweep button. Set the
start and finish frequencies that you
wish to conduct your sweep across
these can be inputted using the
numerical buttons on the PSM’s front
panel. Set the amount of steps (Data
Points) within your sweep to be
displayed within the results screens.