Command:
GRAT1LINES
Purpose:
Set the line spacing for grating 1 in lines per mm.
Parameter:
An integer number from 0 (mirror) to 4000.
Example:
grat1lines 600
: set the lines/mm for grating 1 to 600.
Default:
600.
Command:
GRAT1LINES?
Purpose:
Query grating 1’s line spacing setting.
Example:
grat1lines?
E.g. response
1200
: the number of lines per mm for grating 1 has been set to 1200.
Command:
GRAT1OFFSET
Purpose:
Set the calibration offset for grating 1.
Parameter:
A floating point number.
Example:
grat1offset -2.1
: set the cal offset for grating 1 to -2.1.
Default:
0.0.
Command:
GRAT1OFFSET?
Purpose:
Query grating 1’s line calibration offset setting.
Example:
grat1offset?
E.g. response
1.5
: the cal offset for grating 1 has been set to 1.5.
To set the settings for the other gratings using the previous grat1nnn commands, replace the
number “1” with the desired grating number.
Command:
HANDSHAKE
Purpose:
Set the handshake setting. See the “Handshake Mode” section.
Parameter:
0 to turn off handshake, 1 to turn on.
Example:
handshake 1
: turn on handshake mode
Default:
0 (disabled)
Command:
HANDSHAKE?
Purpose:
Query the handshake setting.
Example:
Handshake?
E.g. response
1
: handshake mode is turned on
Содержание Cornerstone 130B
Страница 1: ...90088262MCS130B Rev B s 1 8m Monochromator Family User s Manual Cornerstone 130B...
Страница 32: ...90088262 MCS130B CORNERSTONE 130B MONOCHROMATORS 32 Figure 23 Model 74010 Filter Wheel...
Страница 42: ...90088262 MCS130B CORNERSTONE 130B MONOCHROMATORS 42 14 13BDIMENSIONS Figure 30 Cornerstone 130B Dimensions...
Страница 44: ...90088262 MCS130B CORNERSTONE 130B MONOCHROMATORS 44 Figure 33 Model 74006 Mounting Plate...